Precision You Can Trust
NABL-Accredited Calibration | Available on Request
The Pico Technology TA345 is a standard-grade manual SOLT calibration kit for two-port vector network analyser calibration from 300 kHz to 8.5 GHz. It combines 50 Ω SMA female short, open, load and through standards in a four-piece, five-port assembly. Serial-number-specific traceable calibration data, tight-tolerance stainless steel interfaces and a protective carrying case support reliable VNA calibration for RF development, production testing, education and general laboratory measurements..
The Standard 8.5 GHz SOLT Calibration Kit with SMA Female Connectors, order code TA345, is a manual mechanical calibration kit designed for PicoVNA and compatible 50 Ω vector network analysers.
The kit contains:
The standards are supplied as an assembled five-port unit. They can be separated for individual use, connector inspection, cleaning or refurbishment if one calibration component becomes damaged.
SOLT calibration uses four known RF reference conditions:
During calibration, the operator manually connects each standard when requested by the VNA software.
The resulting error correction can compensate for systematic measurement errors such as:
The TA345 is not an automated E-Cal module. The RevineTech page should remove claims related to electronic switching, USB operation and one-touch calibration.
The TA345 uses SMA female connectors across its short, open, load and through standards.
It is suitable for calibration setups terminating in:
The corresponding standard-grade male kit is the TA344.
The TA345 should not be described as a mixed male-and-female kit. All five connection ports in the assembled calibration kit are SMA female.
The TA345 is characterised from:
This supports VNA calibration for measurements involving:
The page should use the complete range of 300 kHz to 8.5 GHz, rather than only stating 8.5 GHz.
Pico Technology describes the TA345 as a four-piece, five-port SOLT calibration kit.
The assembly contains:
Together, these provide five SMA female connection ports.
This construction allows the standards to be:
The modular format is useful for laboratories where calibration standards are used frequently.
Each TA345 kit is supplied with traceable calibration data linked to its individual serial number.
The data characterises the actual frequency-dependent behaviour of the:
This allows the VNA to account for non-ideal properties such as:
The standards should therefore be used with the calibration data supplied for that specific kit, rather than with generic or ideal standard values.
Pico Technology offers standard and premium grades of its manual SOLT kits.
The TA345 is the standard-grade SMA female version and is suitable for regular working calibration, production testing and general laboratory measurement.
| Feature | TA345 Standard Kit | TA347 Premium Kit |
|---|---|---|
| Connector type | SMA female | PC3.5 female |
| Grade | Standard | Premium |
| Frequency range | 300 kHz to 8.5 GHz | 300 kHz to 8.5 GHz |
| Calibration type | Manual SOLT | Manual SOLT |
| Construction | Four-piece, five-port | Four-piece, five-port |
| Characterisation | Serial-number-specific | Serial-number-specific |
| Typical use | Working and transfer standard | Precision or reference calibration |
Many laboratories use:
Pico specifically notes that customers may retain both premium-grade and standard-grade calibration standards for different measurement roles.
The TA345 load standard provides the following published return-loss performance:
| Frequency Range | Uncorrected Return Loss | Corrected Return Loss |
|---|---|---|
| Up to 3 GHz | ≥30 dB | ≥40 dB |
| Above 3 GHz to 8.5 GHz | ≥26 dB | ≥37 dB |
Corrected return loss applies when the supplied serial-number-specific calibration data is used.
The load standard supports correction of:
The open standard provides the following published return-loss limits:
| Frequency Range | Return Loss |
|---|---|
| Up to 3 GHz | ≤0.15 dB |
| Above 3 GHz to 8.5 GHz | ≤0.20 dB |
The open is not treated as an ideal infinite impedance. Its frequency-dependent capacitance, delay and connector effects are included in the supplied characterisation.
The short standard provides:
| Frequency Range | Return Loss |
|---|---|
| Up to 3 GHz | ≤0.20 dB |
| Above 3 GHz to 8.5 GHz | ≤0.25 dB |
The calibration data accounts for the standard’s actual delay, loss and inductive characteristics.
The through adaptor provides:
The through connection is used to calculate forward and reverse transmission-tracking and load-match correction terms during two-port calibration.
A two-port SOLT calibration normally includes:
The calibration kit improves systematic error correction, but measurement quality still depends on the cables, connectors, torque, environment and device under test.
SMA calibration standards must be handled carefully.
Recommended practices include:
Connector wear or contamination can increase measurement uncertainty even when the correct calibration data is used.
After calibration, an independent check standard can help verify the complete measurement setup.
Compatible Pico check standards include:
Verification can help identify:
The SOLT kit creates the VNA error correction. A check standard tests whether the calibrated system produces the expected result.
The TA345 can be used with:
The selected sweep range must remain within the kit’s 300 kHz to 8.5 GHz characterisation.
The Pico TA345 is suitable for:
Includes mechanical:
Supports VNA calibration across RF, wireless and microwave applications.
Corrected return loss:
Supports secure storage and transport of the calibration standards.
| Specification | Pico TA345 |
|---|---|
| Product name | Standard 8.5 GHz SOLT Calibration Kit, SMA Female |
| Order code | TA345 |
| Product type | Standard-grade manual mechanical VNA calibration kit |
| Calibration method | Two-port SOLT |
| Standards included | Short, open, load and through |
| Construction | Four-piece, five-port assembly |
| Connector type | SMA female |
| Nominal impedance | 50 Ω |
| Characterised frequency range | 300 kHz to 8.5 GHz |
| Connector construction | Tight-tolerance stainless steel |
| Characterisation | Serial-number-specific traceable calibration data |
| Electronic switching | No |
| USB interface | No |
| USB power | No |
| Internal temperature oven | No |
| Carrying case | Included |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Uncorrected return loss | ≥30 dB | ≥26 dB |
| Corrected return loss | ≥40 dB | ≥37 dB |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Return loss | ≤0.15 dB | ≤0.20 dB |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Return loss | ≤0.20 dB | ≤0.25 dB |
| Specification | Value |
|---|---|
| Maximum insertion loss | 0.25 dB |
| Maximum characterised frequency | 8.5 GHz |
| Connector configuration | SMA female to SMA female |
| Nominal impedance | 50 Ω |
| Characterisation data | Included |
| Specification | Details |
|---|---|
| Data type | Serial-number-specific calibration characterisation |
| Frequency range | 300 kHz to 8.5 GHz |
| Standards characterised | Short, open, load and through |
| Traceability | Traceable calibration data supplied |
| Suitable use | PicoVNA and compatible VNA calibration workflows |
| Electronic data loading | VNA/software dependent |