Precision You Can Trust
NABL-Accredited Calibration | Available on Request
The Pico Technology CC047 service provides factory recalibration for an existing standard-grade SOLT-STD-M or SOLT-STD-F vector network analyser calibration kit. It applies to Pico TA344 SMA male and TA345 SMA female SOLT standards. The returned kit is recalibrated against traceable references, and updated serial-number-linked calibration data is supplied on a USB memory stick for use in compatible PicoVNA and VNA calibration workflows..
The SOLT-STD-M or SOLT-STD-F Calibration Data Service, product code CC047, is intended for customers who already own a Pico Technology standard-grade manual SOLT calibration kit and require updated traceable calibration data.
The service applies to:
The existing physical calibration kit must be returned to Pico Technology using an authorised Return Material Authorisation number. After recalibration, updated calibration data is supplied on a USB memory stick. Pico states that it only provides this service for its own SOLT calibration standards.
CC047 is a service for an existing calibration kit. It does not include a new set of standards.
The service is intended to update the characterised data for the returned:
Each Pico SOLT kit is originally supplied with traceable calibration data linked to its serial number. CC047 provides recalibration when updated data is required for an existing male or female standard-grade SOLT kit.
| Kit | Order Code | Connector Configuration |
|---|---|---|
| SOLT-STD-M | TA344 | SMA male |
| SOLT-STD-F | TA345 | SMA female |
Both kits are standard-grade, four-piece, five-port manual SOLT calibration assemblies designed for VNA calibration up to 8.5 GHz.
The recalibration service covers the mechanical standards within the applicable kit:
These standards are used during manual two-port SOLT calibration to calculate corrections for systematic VNA measurement errors, including:
CC047 does not convert the kit into an automated E-Cal module. The recalibrated TA344 or TA345 remains a manually connected mechanical SOLT calibration kit.
Pico SOLT calibration kits are supplied with calibration data linked to the individual kit serial number.
This data represents the actual frequency-dependent behaviour of the:
Using kit-specific data allows compatible VNA software to compensate for non-ideal characteristics such as:
CC047 updates this characterisation for the returned standard-grade kit.
After the service is completed, Pico supplies the updated calibration data on a USB memory stick.
The data may be used for:
The USB memory device contains calibration data. It is not used to power or electronically control the SOLT standards.
Each new Pico standard SOLT kit is supplied with traceable calibration data linked to its serial number. CC047 is the corresponding recalibration option for existing standard male or female kits.
Traceable recalibration is relevant for:
The official CC047 listing refers to traceable calibration data. A separate NABL certificate should only be mentioned if RevineTech confirms that it is supplied independently for the specific returned kit.
Pico requires customers to contact support and obtain an RMA number before returning the calibration kit.
A typical return process includes:
The entire matched kit should be returned unless Pico provides different instructions for a particular service case.
The TA344 is the standard-grade SMA male SOLT kit.
It includes:
Its published performance includes corrected load return loss of at least 40 dB up to 3 GHz and at least 37 dB above 3 GHz to 8.5 GHz.
The TA345 is the standard-grade SMA female SOLT kit.
It includes:
Its published corrected load return loss is at least 40 dB up to 3 GHz and at least 37 dB above 3 GHz to 8.5 GHz.
Recalibration may be considered when:
The appropriate interval depends on usage frequency, connector handling, environmental control and the organisation’s measurement assurance policy.
Recalibration updates the characterised data, but it may not correct physical damage.
Before returning the kit, check for:
Pico’s standard kits can be disassembled for individual use or refurbishment if a calibration standard is damaged.
CC047 is suitable for standard SOLT kits used in:
Applies to:
Updated calibration data is supplied on a USB memory stick.
The service does not add:
| Specification | Details |
|---|---|
| Service name | SOLT-STD-M or SOLT-STD-F Calibration Data |
| Pico product code | CC047 |
| Service type | Factory recalibration and updated calibration data |
| Applicable male kit | TA344 SOLT-STD-M |
| Applicable female kit | TA345 SOLT-STD-F |
| Existing calibration kit required | Yes |
| Return to Pico required | Yes |
| RMA required | Yes |
| New SOLT kit included | No |
| Calibration data supplied | Yes |
| Data-delivery medium | USB memory stick |
| Kit ownership | Pico provides calibration only for its own SOLT standards |
| Calibration-kit type | Standard-grade manual mechanical SOLT |
| Specification | TA344 | TA345 |
|---|---|---|
| Pico type | SOLT-STD-M | SOLT-STD-F |
| Connector gender | SMA male | SMA female |
| Nominal impedance | 50 Ω | 50 Ω |
| Standards | Short, open, load, through | Short, open, load, through |
| Construction | Four-piece, five-port | Four-piece, five-port |
| Maximum characterised frequency | 8.5 GHz | 8.5 GHz |
| Operation | Manual | Manual |