The PicoVNA TA519 is a USB-controlled automated electronic calibration module for two-port vector network analyser calibration. It covers 300 kHz to 8.5 GHz and provides two 50 Ω SMA female ports with internally switched short, open, load and through standards. Embedded full S-parameter characterisation, oven-controlled temperature stabilisation and PicoVNA software integration support fast, repeatable SOLT calibration with fewer manual RF connections..
The PicoVNA E-Cal 8.5 GHz Automated Calibration Kit with SMA Female Ports, order code TA519, is designed to simplify two-port VNA calibration while reducing repeated connection and disconnection of individual mechanical standards.
The module combines:
The TA519 is suitable for laboratories, production facilities, universities, RF development teams and automated test systems that regularly perform two-port vector network analyser calibration.
SOLT calibration uses four standards:
A conventional manual calibration requires each standard to be connected separately during the calibration process.
This can involve:
Every additional connection can introduce:
The TA519 reduces the manual process to one RF connection. PicoVNA software then automatically switches between the module’s internal calibration states and reads its stored characterisation data through USB.
The TA519 provides:
The female configuration is suitable where the VNA test setup terminates in SMA male connectors.
Typical compatible configurations include:
The TA519 should not be described as having both male and female ports. Both RF calibration ports are SMA female.
For SMA male calibration ports, the corresponding Pico model is the TA518 SOLT-AUTO-M.
The TA519 is fully characterised from:
This range supports VNA calibration for applications involving:
The page should use the full specification of 300 kHz to 8.5 GHz, rather than only stating 8.5 GHz.
The TA519 electronically switches between internally incorporated:
These states allow PicoVNA software to calculate correction terms for systematic measurement errors.
A two-port SOLT calibration can correct for:
Once calibration is completed, the measurement reference planes are established at the selected test-port or cable endpoints.
Electronic calibration standards are not perfectly ideal.
Internal RF switches, connectors and transmission paths introduce small differences between the actual module response and a theoretical open, short, load or through.
Pico Technology addresses this by supplying:
PicoVNA software uses the actual characterised response of the module during calibration rather than assuming mathematically ideal standards.
The module stores its characterisation records internally.
When connected through USB, PicoVNA software can:
This eliminates the need to manually locate and load separate standard definition files for every calibration operation.
The TA519 includes a fast-response internal oven that stabilises the calibration standards.
Published environmental performance includes:
Temperature stabilisation helps reduce changes in the electrical response of the module caused by ambient temperature variation.
This supports:
The module should be allowed to complete its warm-up period before calibration begins.
The TA519 uses a USB 2.0 Micro interface for:
A separate power adaptor is not required for the E-Cal module.
The USB connection controls the calibration module but does not carry the RF measurement signal. RF calibration takes place through the two SMA female ports.
The TA519 provides published residual calibration performance of:
These figures describe the calibration performance associated with the characterised module under the specified operating conditions.
Final VNA measurement accuracy also depends on:
The TA519 RF ports have the following published limits:
The protection limit should not be treated as a normal operating level.
Before calibration, the VNA test-port power should be checked to ensure it remains within the module’s specified operating range.
Pico Technology states that its VNA calibration and check standards are calibrated against fully traceable PC3.5 standards.
Traceability is valuable for:
The product page should describe the module as having traceable characterisation data.
An NABL calibration certificate should only be mentioned when RevineTech has separately confirmed that one is included in its commercial package.
A typical TA519 calibration workflow includes:
Automated switching reduces manual operations, but correct connector handling and cable positioning remain important.
Traditional mechanical calibration may require several standards to be connected and disconnected during each calibration.
Automated E-Cal can reduce:
This is particularly useful in production facilities and laboratories where calibration is repeated frequently.
The TA519 helps improve repeatability by combining:
More frequent and repeatable calibration can help maintain measurement confidence when fixtures, cables or environmental conditions change.
The TA519 is suitable for automated and semi-automated RF measurement environments.
Applications include:
The calibration module can be incorporated into a controlled test procedure using PicoVNA software and compatible automation workflows.
The 300 kHz to 8.5 GHz frequency coverage supports calibration of current PicoVNA instruments operating within this range.
Typical use includes:
The selected VNA sweep must remain within the characterised bandwidth of the E-Cal module.
| Feature | TA519 | TA518 |
|---|---|---|
| Product type | Automated E-Cal module | Automated E-Cal module |
| RF ports | Two SMA female | Two SMA male |
| Frequency range | 300 kHz to 8.5 GHz | 300 kHz to 8.5 GHz |
| Calibration method | Two-port SOLT | Two-port SOLT |
| Power and control | USB | USB |
| Temperature control | Internal oven | Internal oven |
| Typical use | Male test-cable ends | Female test-cable ends |
Users requiring both connector genders can consider the TA520 Dual E-Cal Kit, which contains TA518 and TA519 modules.
The PicoVNA TA519 is suitable for:
Internally switched standards include:
The module provides:
Supports VNA calibration across low-frequency RF, wireless and microwave applications.
USB provides:
The module is characterised against traceable PC3.5 reference standards.
| Specification | PicoVNA TA519 |
|---|---|
| Product name | PicoVNA E-Cal 8.5 GHz Automated Calibration Kit, SMA Female |
| Pico order code | TA519 |
| Type | SOLT-AUTO-F |
| Product type | Ovened USB-controlled automated E-Cal SOLT module |
| Calibration method | Two-port SOLT |
| Internal standards | Short, open, load and through |
| Additional standard | Separately characterised and polarised port adaptor |
| RF ports | 2 |
| Connector type | SMA female |
| Port impedance | 50 Ω |
| Characterised frequency range | 300 kHz to 8.5 GHz |
| Characterisation | Full multipath S-parameter data |
| Data storage | Embedded internal memory |
| Control and power | USB 2.0 Micro |
| Temperature control | Internal oven |
| Software | PicoVNA software |
| Parameter | Published Value |
|---|---|
| Directivity | 40 dB |
| Source match | 40 dB |
| Load match | 36 dB |
| Reflection tracking | 0.05 dB |
| Transmission tracking | 0.04 dB |
| Transfer calibration method | SOLT comparison |
| Specification | Value |
|---|---|
| Port count | 2 |
| Port impedance | 50 Ω |
| Connector | SMA female |
| Normal operating input limit | +10 dBm |
| Protection limit | +20 dBm or 1 V peak |
| Bandwidth | 300 kHz to 8.5 GHz |
| Specification | Details |
|---|---|
| USB interface | USB 2.0 Micro |
| USB function | Power, control and data |
| Characterisation data | Read automatically from module |
| Internal switching | Controlled by PicoVNA software |
| Separate power adaptor | Not required |
| Compatible software | PicoVNA |
| Specification | Details |
|---|---|
| Operating temperature | 5°C to 40°C |
| Oven-control range | 18°C to 28°C for quoted accuracy |
| Typical oven warm-up | 45 seconds at 23°C |
| Operating humidity | 5% to 80% RH, non-condensing |
| Storage temperature | −20°C to 50°C |
| Storage humidity | 5% to 80% RH, non-condensing |
| Specification | Details |
|---|---|
| Length | 65 mm |
| Width | 43 mm |
| Height | 15 mm |
| Dimensions include connectors | Yes |
| Weight | Approximately 60 g |
| Enclosure | Compact portable calibration module |