Precision You Can Trust
NABL-Accredited Calibration | Available on Request
The Pico Technology CC050 service provides factory recalibration for an existing CHK-NON-F or CHK-INS-MF vector network analyser check standard. The returned standard is recharacterised, and updated Touchstone S-parameter measurement data is supplied on USB memory. The reference data is traceable through PC3.5 standards to national standards and can be used to validate the calibration and measurement accuracy of PicoVNA or other compatible VNAs..
The CHK-NON-F or CHK-INS-MF Check Standard Recalibration Service, product code CC050, is intended for customers who already own one of Pico Technology’s RF check standards and require updated traceable reference data.
The service applies to:
The existing physical check standard must be returned to Pico Technology using an authorised Return Material Authorisation number. After recalibration, updated measurement data is supplied on a USB memory device.
A check standard is a stable and fully characterised RF device used to verify whether a vector network analyser measurement setup and its software calibration are producing reliable results.
It is measured after completing the VNA calibration. The live S-parameter results are then compared with the traceable reference data supplied for the standard.
This process can help identify:
A check standard validates the measurement setup. It is not itself used to perform the initial SOLT, TRL, TRM or electronic calibration of a VNA.
CC050 is specifically intended for recalibrating an existing TA430 or TA431 check standard.
The service includes:
It does not include:
Customers purchasing a new check standard should order TA430 or TA431 separately.
The recalibration data is supplied in the industry-standard Touchstone format.
Touchstone files contain characterised S-parameter data that can be imported into compatible VNA software for comparison with a new live measurement.
The reference data supports verification of:
Because the files use an industry-standard format, they can also be used to manually validate measurements made with many other manufacturers’ vector network analysers.
The supplied measurement data is traceable through PC3.5 reference standards to national standards.
Traceability provides a documented measurement chain linking the characterised check standard to recognised measurement references.
This is valuable for:
The page should describe this as traceable measurement data rather than automatically claiming that a separate printed or NABL-accredited calibration certificate is included. Pico’s official CC050 listing specifies Touchstone data on USB memory but does not state that an NABL certificate is part of this service.
The TA430 CHK-INS-MF has:
It supports full S-parameter validation following an insertable 12-term VNA calibration.
The TA431 CHK-NON-F has:
It supports full S-parameter validation for non-insertable, known-through and unknown-through calibration configurations.
Both supported check standards use a short length of deliberately mismatched transmission line.
The device produces a predictable, smooth and stable reflection and transmission response across its characterised frequency range.
This makes it more demanding and informative than validating a VNA using only a well-matched through connection. It tests whether the calibrated system can accurately measure a device with a significant and changing mismatch.
Recalibration may be considered when:
Recalibration does not automatically repair physical connector damage. The condition of the returned device should be assessed before shipment.
A typical validation process includes:
PicoVNA software includes comparison functionality for evaluating the measured result against the characterised check-standard data and relevant measurement uncertainties.
The recalibrated check standard can be used with:
The TA430 and TA431 standards are characterised up to 8.5 GHz, covering the operating frequency ranges of both current PicoVNA models.
Before sending the check standard:
Pico specifically instructs customers to contact support for an RMA number before returning the check standard.
The CC050 recalibration service is suitable for check standards used in:
Both supported check standards are characterised from:
Reference data supports comparison of:
The Touchstone reference files can be used to validate compatible vector network analysers from different manufacturers.
An RMA number must be obtained before the existing check standard is returned to Pico Technology.
| Specification | Details |
|---|---|
| Service name | CHK-NON-F or CHK-INS-MF Check Standard Recalibration |
| Pico product code | CC050 |
| Service type | Factory recalibration and recharacterisation |
| Applicable products | TA430 and TA431 |
| Physical product included | No |
| Existing standard required | Yes |
| Return to Pico required | Yes |
| RMA required | Yes |
| Supplied result | Updated Touchstone measurement data |
| Data-delivery medium | USB memory |
| Traceability | Via PC3.5 standards to national standards |
| Compatible instruments | PicoVNA and other Touchstone-compatible VNAs |
| Model | Product | Connector Configuration | Calibration Setup |
|---|---|---|---|
| TA430 | CHK-INS-MF | SMA male to SMA female | Insertable |
| TA431 | CHK-NON-F | SMA female to SMA female | Non-insertable |
| Specification | TA430 / TA431 |
|---|---|
| Characterised frequency range | 300 kHz to 8.5 GHz |
| Nominal line impedance | 25 Ω |
| Mismatched-line length | 75 mm |
| S-parameters | S11, S21, S12 and S22 |
| Reference-data format | Touchstone |
| Typical purpose | Verification of VNA setup and calibration |
| Compatible PicoVNAs | PicoVNA 106 and PicoVNA 108 |
| Check Standard | Port 1 | Port 2 |
|---|---|---|
| CHK-INS-MF, TA430 | SMA female | SMA male |
| CHK-NON-F, TA431 | SMA female | SMA female |